ResQGARD

First Response Australia
ResQGARD  (Impedance Threshold Device)

The ResQGARD is an Impedance Threshold Device (ITD) that provides a safe, simple and convenient way to treat states of low blood pressure in spontaneously breathing patients.

Animal and clinical studies have shown that the ResQGARD increases blood pressure during hypotension from a variety of causes, for example:

  • Orthostatic intolerance
  • Hypovolemia
  • Heat shock
  • Dialysis
  • Blood donation

The ResQGARD provides therapeutic benefit as soon as a patient begins to breathe through it.

The ResQGARD can be used on a facemask or with a mouthpiece.



Using the ResQGARD with a Facemask

1. Connect the ResQGARD to the facemask.
2. Explain to the patient that they will feel some slight resistance when
inhaling. This means that the device is working and helping to improve blood flow.
3. Hold the mask over the nose and mouth maintaining a tight facemask seal.
4. A ResQStrap can be used to hold the ResQGARD in place.
5. Breathe at 10-16/min.
6. Inhale slowly (over 2-3 secs) and deeply; exhale normally.
7. If supplemental oxygen is desired, connect the oxygen tubing to the port and deliver up to 15 lpm.


Using the ResQGARD with a Mouthpiece

1. Connect the ResQGARD to the mouthpiece.
2. Explain to the patient that they will feel some slight resistance when
inhaling. This means that the device is working and helping to
improve blood flow.
3. Place the mouthpiece into the mouth and maintain a tight seal with
the lips.
4. Breathe through the mouth only @ 10-16/min. An optional nose clip
prevents breathing through the nose.
5. Inhale slowly (over 2-3 secs) and deeply; exhale normally.
6. If supplemental oxygen is desired, connect the oxygen tubing to the
port and deliver up to 15 lpm.


How It Works

During inspiration, a negative pressure (created
from expansion of the thorax) draws air into the
lungs. When inspiratory impedance is applied to
the breathing circuit, it enhances the negative
pressure (vacuum) in the chest, which pulls more
blood back to the heart, resulting in increased
preload and thus, enhanced cardiac output on
the subsequent cardiac contraction.

How It Works VIDEO CLIP (wmv file 12megs)

Technical Specifications
Construction Description: Material
Exterior Housing: Polycarbonate
Interior Molded Components: Polycarbonate
Diaphragm: Silicone
Valve Gasket: Silicone
Valve Spring: Nickel coated stainless steel

Physical Specifications
Patient Port:
15 millimeters ID
22 millimeters OD

ISO Standard Supplemental Oxygen Port
Height: 7 centimeters (2.76 inches)
Diameter: 5.2 centimeters (2.0 inches)
Weight: 32 grams
Dead Space: 14 milliliters
Valve Cracking Pressure: -7 cmH2O
Exhalation Impedance:< -1.2 cmH2O
Operating Conditions: -18 degrees to +45 degrees C.
Storage Conditions: -40 degrees to +60 degrees C.
Shelf Life: Four years

Accessories:
Facemask: High quality facemask with three exhalation ports
Mouthpiece: 22 mm ID
Nose Clip: Foam tip clip, EVA/POM
ResQStrap: Adjustable head strap

 


PRODUCT INFORMATION SHEET

ResQGARD Summary click here